Allows to analyze the content of hydrogen sulfide and carbon dioxide in the recovered and non-regenerated amine in order to improve the efficiency of the desulfurization process.
- MEMS technologies
Purpose
Thechnical basis of this analyzer is prrocess GC PGC 90.50 with additional vapor phase analysis. Supplied togeather with sample preparation system, frame for suppor gases cylinders. Sample preparation system could ave valves for switching samples from different lines.
Includes our own software Analizator. Using it operator is able to receive all nesessary data, see the chromatograms, make configuration and parametrization. Both English and Russian languages are available
Features
Compared to optical analyzers this development has the possibility to control the concentration of CO2, operating with По сравнению с оптическими анализаторами у данной разработки имеется возможность контроля содержания углекислого газа, а также возможность работы с turbid mixtures.
Specification:
Measuring range of H2S |
0.02 – 10.0 g/l |
Measuring range of CO2 |
0.02 – 10.0 g/l |
Explosion protection |
EExdIICT4 |
Type of detector |
ДТП |
Temperature range |
0-40С |
Carrier gas |
Helium А (99.995%) |
Power |
55 W |
Calibration |
Automatic by calibration gas |
Detection limit |
0.002 g/l |
Data transfer |
RS485, Ethernet |